12.10

Measurement

12.10.1

Selecting the Measuring Program

MEASMODE EDGEHL|EDGELH|DIA|GAP|SEGMENT
Defines the measuring program.
EDGEHL: Light-dark edge; position of first light-dark edge
EDGELH: Dark-light edge; position of the first dark-light edge
DIA: Diameter (distance from first light-dark edge to dark-light edge) and position of the diameter center axis
GAP: Gap (distance from first dark-light edge to light-dark edge) and position of the gap center axis
SEGMENT: Segment (difference between the two selected edges and positions of the various center axes)
See DEFSEG1...DEFSEG8
The position is the distance to the beginning of the line (see MEASDIR).
12.10.2

Edge Search Direction

SEARCHDIR STANDARD|INVERSE
Sets the edge search direction.
STANDARD: Searches for edges by moving away from the mounting rail (the first edge is the one next to the mounting
rail)
INVERSE: Searches for edges by moving toward the mounting rail (the first edge is the one furthest away from the
mounting rail)
12.10.3

Measurement Direction

MEASDIR STANDARD|INVERSE
Sets the measurement direction for the edges.
STANDARD: The measurement direction begins at the start of the measuring range (SMR, mounting rail) and moves
toward the end of the measuring range.
INVERSE: The measurement direction begins at the end of the measuring range (EMR) and moves toward the start of
the measuring range (mounting rail).
12.10.4

Number of Expected Edges

EXPEDGES <value>
Defines the number of expected edges for the ERROROUT command.
Value: 0 ... 251