12.14.2
Possible Signals for Error Outputs
META_ERRORLIMITSIGNAL1
META_ERRORLIMITSIGNAL2
META_ERRORLIMITSIGNAL3
Lists all the available signals that can be selected for the corresponding ERRORLIMITSIGNAL1, ERRORLIMITSIGNAL2,
and ERRORLIMITSIGNAL3 commands.
12.14.3
Assigning the Error Output Signal
ERRORLIMITSIGNAL1 [<signal>]
ERRORLIMITSIGNAL2 [<signal>]
ERRORLIMITSIGNAL3 [<signal>]
The command defines or displays the signal for the corresponding switching output. This setting applies to the digital in
puts and outputs (see OUTPUT ERROROUT).
The META_ERRORLIMITSIGNAL1, META_ERRORLIMITSIGNAL2, and META_ERRORLIMITSIGNAL3 commands list
all the available signals of the corresponding switching outputs that can be used for the output.
See also ERRORLIMITCOMPARETO1 and ERRORLIMITVALUES1, ERRORLIMITCOMPARETO2 and ERRORLIMIT
VALUES2, ERRORLIMITCOMPARETO3 and ERRORLIMITVALUES3.
12.14.4
Setting the Upper (Overshooting)/Lower (Undershooting) Limit Value for the Switching Outputs
ERRORLIMITCOMPARETO1 [LOWER|UPPER|BOTH]
ERRORLIMITCOMPARETO2 [LOWER|UPPER|BOTH]
ERRORLIMITCOMPARETO3 [LOWER|UPPER|BOTH]
The command defines or returns the limit value for the corresponding error output.
LOWER: Undershooting
●
UPPER: Overshooting
●
BOTH: Undershooting or overshooting
●
12.14.5
Limits for Overshooting/Undershooting of Switching Outputs
ERRORLIMITVALUES1 [<lower limit [mm]> <upper limit [mm]>]
ERRORLIMITVALUES2 [<lower limit [mm]> <upper limit [mm]>]
ERRORLIMITVALUES3 [<lower limit [mm]> <upper limit [mm]>]
This command defines the lower and upper limit values for the relevant switching output or displays the current setting.
The settings are applied to the relevant digital input/output (see ERROROUT1, ERROROUT2, ERROROUT3).
The ERRORLIMITCOMPARETO1, ERRORLIMITCOMPARETO2, and ERRORLIMITCOMPARETO3 commands define
whether the lower limit value, upper limit value, or both limit values should be applied.
<lower limit [mm]> = -2147.0 ... 2147.0
●
<upper limit [mm]> = -2147.0 ... 2147.0
●