12.11.7
Resetting Statistics
STATISTIC ALL|<signal> RESET
Resets the values for the STATISTICSIGNAL command.
<signal>_MIN
●
<signal>_MAX
●
<signal>_PEAK
●
signal: Measurement data signal (see META_STATISTIC)
●
12.11.8
List of Possible Mastering Signals
META_MASTERSIGNAL
Lists all the available signals that can be selected for the MASTERSIGNAL commands.
12.11.9
Configuring the Master Signal
MASTERSIGNAL [<signal>]
MASTERSIGNAL <signal> <master value>
MASTERSIGNAL <signal> NONE
<signal> A measurement data signal (see META_MASTERSIGNAL)
●
<master value> A value in mm between -2147.0 and 2147.0
●
For displaying, setting, or deleting master entries. The master value is the value with which the current measured value is
aligned if mastering has been activated. Mastering can be activated with the MASTER command. The META_MASTER
SIGNAL command lists all the available signals that can be used for this command. The display output shows all signals
and their currently set master values.
12.11.10
List of Configured Mastering Signals
META_MASTER
Lists the signals (configured with the MASTERSIGNAL command) that can be selected for the MASTER command.
12.11.11
Master Settings
MASTER [<signal>]
MASTER [ALL|<signal> [SET|RESET]]
Defines the mastering settings or displays the current master settings.
The SET master function takes a current measured value from <signal> and <signal> master value (configurable via
MASTERSIGNAL) in order to define the offset. The offset is then applied to all subsequent measured values.
Example: If the master value is 0 and <signal> has a value of 0.5 mm according to the latest measurement, then an offset
of -0.5 mm is applied to <signal>. The RESET function resets the offset to 0. The display output provides a list of signals
and shows the word ACTIVE if mastering is currently active for the signal concerned. If mastering is not active, it shows
the word INACTIVE.
signal: A measurement data signal (see META_MASTER)
●
optoCONTROL 2700
ASCII Communication
Seite 94