6.4.3
Averaging
6.4.3.1
General
Measurement averaging is performed after measured values have been calculated, and before they are issued or proc
essed through the relevant interfaces.
Measurement averaging
improves the resolution,
●
allows the masking of individual interference points, or
●
"smooths" the measurement result.
●
i
Linearity is not affected by averaging. Averaging has no effect on measuring rate and output rate. The
defined type of average value and the number of values must be saved in the sensor to ensure they
are maintained after it has been switched off.
Averaging is disabled by default on the sensor.
Defining Measurement Averaging
►
Switch to the tab
Settings
►
Make the desired settings and confirm them by pressing
optoCONTROL 2700
>
Signal processing
> Calculation.
Save
settings.
Advanced Settings
Seite 57