RECURSIVE: The averaging function uses the previous average value when calculating the new average value. This
permits a high degree of smoothing of the measurement values. A higher <recursive data count> will result in greater
smoothing.
CALC: Used to define custom calculations with one or more signals. The calculation will be performed as follows:
(<factor1> * <signal>) + (<factor2> * <signal>) + <offset> The result of the calculation will be placed inside a new signal
<name>.
THICKNESS: Used to calculate the difference between two signals. This can also be achieved using a CALC calcula­
tion such as: (1 * <signal>) + (-1 * <signal>) + 0
COPY: Used to make a copy of a signal.
NONE: A special option for deleting an entry.
12.11.3

List of Possible Calculation Signals, Computing Module

META_COMP [<id>]
Lists the signals that can be used for the COMP command. The COMP command imposes restrictions concerning which
signals can be used when, and in terms of which signals are permitted.
id: 1 ... 10
12.11.4

Statistic Signals

META_STATISTICSIGNAL
Lists all the signals that are available for the STATISTICSIGNAL command.
12.11.5

Statistics signals settings

STATISTICSIGNAL [<signal>]
STATISTICSIGNAL <signal> NONE|INFINITE|<depth>
Display, configure or delete entries for the statistics settings
Entries for statistics settings generate new signals in the form of:
<signal>_MIN
<signal>_MAX
<signal>_PEAK
These new signals output the minimum value, maximum value and peak value (maximum - minimum) of the last <depth>
measurement cycle of <signal>. The INFINITE option can be selected instead of a fixed depth and causes the new sig­
nals to contain the statistics of all <signal> data. The NONE option is used to delete the statistical configuration. The STA­
TISTIC and RESETSTATISTIC commands are used to reset the values for the new signals. The META_STATISTICSIG­
NAL command lists all available signals that can be used for this command.
Signal: Measurement data signal (see META_STATISTICSIGNAL)
Depth: 2|4|8|...|8192|16384
Number of measured values used to calculate the statistical values of a signal.
12.11.6

Selecting Statistic Signals

META_STATISTIC
Lists all the signals configured with STATISTICSIGNAL that can be selected for the STATISTIC command.
optoCONTROL 2700