12.3.5
Changing the Password
PASSWD <old_password> <new_password> <new_password>
Changes the password for the PROFESSIONAL user level.
Passwords must have at least 1 character and can contain a maximum of 31. The following characters are permitted: a-
zA-Z0-9 _(),;.:-_/.
If the password contains spaces, the entire password must be placed inside quotation marks ("password").
12.4
Correction, Referencing
12.4.1
Light Referencing
LIGHTCORR [ROI] [FORCE]
Calibrates the light correction table (light referencing).
ROI: Light referencing is only performed for the region of interest (defined by ROI command). Areas outside the ROI
●
are set to the factory setting.
FORCE: Overwrites the correction table even if soiling has been detected.
●
12.4.2
Light Correction Status
LIGHTCORRSTATUS
Shows the status of light referencing. Indicates whether or not a light correction is available.
CLEAN: Light correction is available; no soiling was detected within the entire measuring range.
●
WITH_RESTRICTIONS: Light correction is available; no soiling was detected within the ROI.
●
SOILED: Light correction is available; however, soiling was detected during light referencing.
●
NOT_AVAILABLE: No light correction available
●
12.4.3
Printing the Correction Table
LIGHTCORR_PRINT
Prints the light correction table.
12.4.4
Deleting the Correction Table
LIGHTCORR_DEL
Deletes the light correction table.
12.4.5
Soiling Check
SOILINGCHECK [ROI] [OBJ_IGNORE]
Checks the sensor for soiling.
ROI: Only checks the ROI (region of interest, defined by the ROI command)
●
OBJ_IGNORE: Ignores targets (measuring objects); the soiling check is only performed for the exposed sensor area.
●
Completely dark areas with sharp edges are excluded from the check.