Performance Test Record - Agilent Technologies 8510XF Service Quick Reference Manual

Network analyzer system
Hide thumbs Also See for 8510XF:
Table of Contents

Advertisement

Performance Test
Record
Table 4-7
Instrument Model: _________________
Frequency
Minimum Specification
45 MHz
44.999955 MHz
2 MHz
1.999998 GHz
2
20 GHz
19.99998 GHz
26 GHz
26.4999735 GHz
1. The measurement uncertainty is quoted for these performance tests using only the recommended models specified in Table 8-1 of the "Equipment
Required" chapter of the On-Site Service Manual. The quoted uncertainty represents limits of 3 times the equivalent standard deviation (3s) and is
intended to represent 90% confidence level.
2. For 83620A/B or 83621A/B only.
Table 4-8
Instrument Model: _________________
Frequency
Minimum Specification
45 MHz
44.999955 MHz
2 MHz
1.999998 GHz
2
20 GHz
19.99998 GHz
1. The measurement uncertainty is quoted for these performance tests using only the recommended models specified in Table 8-1 of the "Equipment
Required" chapter of the On-Site Service Manual. The quoted uncertainty represents limits of 3 times the equivalent standard deviation (3s) and is
intended to represent 90% confidence level.
2. For 83620A/B or 83621A/B only.
Performance Verification
CW Frequency Accuracy Test
Record the measured frequencies for RF source readings on
for LO source on
Table
4-8.
Performance test record for CW frequency accuracy test (RF source)
Report Number: __________________________
Recorded Results
Performance test record for CW frequency accuracy test (LO source)
Report Number: __________________________
Recorded Results
4-14 8510XF Service Quick Reference Guide
Maximum Specification
45.000045 MHz
2.000002 GHz
20.00002 GHz
26.5000256 GHz
Maximum Specification
45.000045 MHz
2.000002 GHz
20.00002 GHz
Table
4-7, and
Date: _______
1
Uncertainty
10 Hz
10 Hz
4 kHz
5 kHz
Date: _______
1
Uncertainty
10 Hz
10 Hz
4 kHz

Advertisement