Chapter 3: Testing Logic Analyzer Performance

Performance Test Record

Performance Test Record
LOGIC ANALYZER MODEL NO. (circle one): 16753A 16754A 16755A 16756A 16950A
Logic Analyzer Serial No.
Date:
TEST EQUIPMENT USED
Pulse Generator Model No.
Pulse Generator Serial No.
Pulse Generator Calibration Due Date:
MEASUREMENT UNCERTAINTY
Clock Rate
Pulse Generator Frequency Accuracy:
8133A: 1% of setting
Setting: 150 MHz + 2% = 153 MHz
300 MHz + 2% = 306 MHz
TEST RESULTS
Logic Analysis System Self-Tests (Pass/Fail):
Performance Test: Minimum Master to Master Clock Time and Minimum Pulse Width
Pulse Generator Settings
Test Criteria
Pod 1 Results (pass/fail):
Pod 2 Results (pass/fail):
Pod 3 Results (pass/fail):
Pod 4 Results (pass/fail):
60
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
300 Mb/s mode
Freq: 150 MHz plus test margin
Pulse Width: 1 ns less test margin
Test 1 of 2
Test 2 of 2
Eye Finder locates an
Correct number of
eye for each bit
occurrences
detected
Work Order No.
Recommended Test Interval - 2 Year/4000 hours
Recommended next testing:
Oscilloscope Model No.
Oscilloscope Serial No.
Oscilloscope Calibration Due Date:
Pulse Width (Eye Width)
Oscilloscope Horizontal Accuracy:
54845B: ±[((0.007%) (∆t)+(full scale/(2x memory depth))+30
ps] ≅ 30 ps
Oscilloscope Display Resolution: 54845B: ±5 ps
Pulse Width setting: 930 ps
PWmax(worst case): 930 ps+30 ps+5 ps = 965 ps
600 Mb/s mode
Freq: 300 MHz plus test margin
Pulse Width: 1 ns less test margin
Test 1 of 2
Eye Finder locates an
eye for each bit
Test 2 of 2
Correct number of
occurrences
detected