Performance test failure troubleshooting

This section describes the adjustment and troubleshooting procedures used when the
E4980A fails the performance tests. If the performance of the instrument is critical for the
test limits and seems to be adjustable, perform first the adjustment(s) related to the failed
test. When the test result is far from the tolerance of the test or the performance is not
adjustable, isolate the faulty assembly in accordance with the "Performance tests failure
troubleshooting procedure".

Recommended adjustment for performance test failure

Table 4-13 shows the recommended adjustments when the performance test fails. Select
the adjustment program corresponding to the recommended adjustment and perform the
adjustment.
Table 4-13
Recommended adjustment for performance test failure
Test
First failed test
Num.
1
Frequency Accuracy Test
2
DC Source Level Accuracy Test
3
DC Bias Level Accuracy Test
4
DC Bias V Level Monitor Accuracy Test
5
DC Bias I Level Monitor Accuracy Test
6
AC Signal Level Monitor Accuracy Test
7
AC Signal Level Accuracy Test
8
Impedance Measurement Accuracy Test
9
DCR Measurement Accuracy Test
Chapter 4
Performance test failure troubleshooting
Recommended adjustment
A1
A2
Troubleshooting
A3
PPMC
57