Connecting A Bioreactor Vessel; Agitator - Thermo Scientific HyPerforma G3Lite User Manual

Bioreactor controller
Table of Contents

Advertisement

Chapter 2
|
Setup
Table 2.2. Heated vessel power rating.
Thermo 50 L (455 W)
Thermo 100 L (865 W)
Thermo 250 L (1357 W)
Thermo 500 L (2600 W)
Thermo 1000 L (5000 W)
Thermo Scientific

2.8 Connecting a bioreactor vessel

Power through the controller is sufficient to run all the equipment
associated with a 50 L S.U.B. Please ensure that the appropriate
adapter based on voltage from the IEC 60309 (blue) connector to the
6-pin gray connector is in place.
Note: At 120 V, power through the controller is NOT sufficient to run all
the equipment associated with a 100 L S.U.B. If you are using a 100 L
S.U.B. on 120 V, ensure that the S.U.B. has a G3Lite External Heater
Power box rated for the correct wall and heater voltages. For low and
medium power sub configurations, the power can be drawn from
the G3Lite. The G3Lite can be updated to reflect the configuration in
Table 2.2.
At 240 V, power through the controller is sufficient to run all the
equipment associated with a 100 L S.U.B. If you are using a 100 L
S.U.B. on 240 V, ensure that the appropriate adapter from the IEC
60309 (blue) connector to the 6-pin connector is in place.
Vessel size
120 VAC
Low (3.8 A)
Low (7.2 A)
Medium (11.3 A)
N/A
N/A

2.9 Agitator

If using an AC agitator, the connector cable exits the controller from
the upper right side panel. This cable attaches directly to the AC
motor junction box or to a mating cable as appropriate. If the size of
the vessel is different from what the controller is originally configured
for, please contact Thermo Fisher Scientific for reprogramming of the
variable frequency drive (VFD).
Connection for a D Agitator is on the lower end of the main connectors
shown in section 1.4. This is a double connector, with one connector
for the signal and the other for power to control a motor.
Voltage
240 VAC
Low (1.9 A)
Low (3.6 A)
Medium power 120–240 VAC or low power
Low (5.7 A)
Medium (10.8 A)
High (20.8 A)
HyPerforma G3Lite Bioreactor Controller User's Guide
Description
Low power 120–240 VAC
Low power 120–240 VAC
120–240 VAC
Medium power 240 VAC
High power 240 VAC
|
30

Advertisement

Table of Contents
loading

Table of Contents