PROCESS
01
PREPARE
02
CLEAN
PREPARE
20 s - Immediately after use, rinse the lumen(s) line of the handpiece/instrument with water.
Coarse soiling must be removed immediately after application.
Safely transport to the reprocessing area to avoid any damage to the products, contamination of the environment
or of the persons involved in the reprocessing process.
Cleaning is to be performed within 1 hour from use.
Wear personal protective equipment according to the type of preparation.
AIRFLOW
and PERIOFLOW
®
Always carry out handpiece powder unclogging and
check for both lumens (water and powder) clearance
before proceeding. Use Easy Clean.
PIEZON
Endochuck:
®
Remove any installed file.
Remove the screw and separate the small O-ring.
Any product can be cleaned manually or automatically by washer or disinfector.
EMS recommends the use of an ISO 15883 compliant automatic washer-disinfector (WD) for optimal
effectiveness and service life.
AUTOMATIC
CLEAN, DISINFECT AND DRY
The washer-disinfector must comply with ISO 15883,
must have suitable baskets to hold small fragile products
and must have rinsing connections with diameter of
approximately 16mm for the attachment to product
lumen.
Correctly place the product into a suitable rack, connect all
lumens to the rinsing connectors and start the automated
cleaning.
Instructions of the manufacturer of the washer
disinfector must also be observed.
In case of use of a chemical disinfectant, follow
carefully the instructions provided by the disinfection
solution manufacturer.
For example: MediClean Dental, Dr. Weigert Hamburg
1
Medisafe MED100.33
2
03
DISINFECT
04
DRY
Handpieces:
®
Remove any installed instrument.
Remove the nose cap and separate the light guide and O-ring.
Separate the PIEZON
CombiTorque
and Combitorque
05
INSPECT
06
PACK
PIEZON
Handpieces:
®
PIEZON
Instruments with CombiTorque
®
Instrument and the
®
but keep each original instrument
®
together throughout their life.
®
MANUAL
CLEAN
The following process can be used with any EMS
product:
Brush the devices in a cleaning solution of 0.5% neodisher
in deionized water at 40°C with a suitable soft bristled
brush
until all visible residue has been removed.
2
For products with lumens, flush all lumens
15 s
with a spray gun (water jet gun, with a static
water pressure of 2 bar) with cold tap water.
Place the product in 0.5% solution of
15 min
neodisher
with deionized water at 40°C.
1
Make certain that all lumens are filled with
cleaning solution (use a syringe if needed).
Rinse all lumens by flushing with a spray gun
15 s
(water jet gun, with a static water pressure
of 2 bar) with cold deionized water.
Rinse the whole product under cold running
10 s
deionized water.
07
STERILIZE
08
STORE
:
®
1