BEFORE USE
Please read these Operating Instructions carefully as they
explain all the most important details and procedures. Please
pay special attention to the safety precautions.
Always keep these Instructions close at hand.
These Instructions are only applicable to the equipment
they were delivered with.
DEVICES AND HANDPIECE CORDS
The following instructions have been validated for all reusable products
the processor to ensure that the processing, as actually performed using the equipment, materials and personnel in the processing
facility, achieves the desired result. This requires validation and/or validation and routine monitoring of the process. Likewise any
deviation by the processor from the instructions provided should be properly evaluated for effectiveness and potential adverse effects.
AIRFLOW
and PIEZON
®
INTRODUCTION
EMS recommends compliance with cleaning, disinfection,
packaging for sterilization and sterilization procedures according
to ISO 17664.
Always report adverse events related to device reprocessing
directly to EMS.
Products must be cleaned, disinfected and, if applicable,
sterilized prior to first use. Do not reprocess the products over
the authorized number of sterilization cycles. To replace, refer
to the "Service life" section of these Instructions.
Users shall also observe any legal requirements applicable in their country as well as the hygiene regulations in effect
at the hospital or clinic.
This applies especially with regard to additional requirements for the inactivation of prions.
SERVICE LIFE
The products have been designed for a large number of
sterilization cycles. The materials used in their manufacture
were selected accordingly. However, with every renewed
preparation for use, thermal and chemical stresses will result
in ageing of the products.
Lifetime : see product's Instructions for Use.
SYMBOLS
Caution / Danger
Risk of property damage or injury
Please note
Useful additional information and hints
Prohibited
Wear personal protective equipment
Hereinafter referred to as "products".
1
4
devices
Handy
®
To prevent injury to persons or damage to property, please
observe the corresponding directives and symbols.
This instruction follows the recommendations of RKI - Robert
Koch Institut, Germany.
Translations in this manual are based on the English version.
Thus the English text will be referred to in case of discrepancy.
of EMS dental devices. It remains the responsibility of
1
Wipe the surface with a wet cloth until all visible
contamination is removed.
DISINFECT
Wipe the surface with 70% isopropyl alcohol or a
disinfection wipe with proven efficacy.
PIEZON
Kit on your unit
®
Do not use CaviWipes™.
Concentrations and contact times specified by the
manufacturer of the cleaning and disinfection agent must be
observed.
Remember that sterilization cannot be achieved unless
assembly components are cleaned and disinfected first.
If parts of these instructions are not clear or seem to be
inadequate, do not hesitate to contact/inform EMS.
Always replace products that present signs of wear or
early degradation, regardless of the number of sterilization
cycles left unused.
DO NOT expose the products to temperature exceeding
138°C.
Refer to instruction manual
Sterilizable at up to 135°C in the autoclave
Suitable for thermal disinfection
Required time
Abbreviations used: min (minute) ; s (second)
CLEAN
Follow carefully the instructions provided by the
disinfection solution manufacturer.
It damages EMS products.