The following automated process
achieve an A0 Level of 3000:
2 min Pre-clean with cold tap water
Drain
5 min Clean at 55°C with tap water
cleaning solution.
Drain
3 min Rinse and neutralize with cold
demineralized water
Drain
2 min Rinse with cold demineralized water
Drain
3 min Perform thermal disinfection (final rinse)
with deionized water at 93°C minimum.
Drain
20 min
Dry at 100°C.
(at least)
INSPECT
If stains are still visible on the product after cleaning/disinfection, the entire cleaning/disinfection procedure
must be repeated. Products with visible damage, chip/flake loss, corrosion or deformation must be disposed of (no
further use is allowed). Check also the integrity of O-rings and gaskets and replace if damaged or deformed.
Verify the product to be fully dry. In case of detection of residues of water, remove these using an air pistol (clean
compressed air). Fully dry the lumen and the whole product until no more residues of water are present (visible or
detectable).
For example: Miele Professional G 7836 CD with Miele Rack E429
1
6
cold tap water = 16°C +/- 2°C
2
cold demineralized water = 20°C +/- 2°C
3
AUTOMATIC
can be used to
1
.
2
and 0.5%
2
.
3
DO NOT use any ultrasound bath cleaning
procedure with the handpieces: it may destroy the
products.
The following process is intended ONLY for EMS
PIEZON
Instruments and file holders:
®
For product with lumens, flush all lumens
15 s
with a spray gun (water jet gun, with a static
water pressure of no more than 2 bar) with
cold tap water.
10 min Place products in a metal sieve and sonicate
in 0.5% cleaning solution of neodisher
.
3
deionized water at 40°C
all lumens are filled with cleaning solution
(use a syringe if needed).
Rinse all lumens by flushing with a spray
15 s
gun (water jet gun, with a static water
pressure of no more than 2 bar) with cold
deionized water.
Rinse the whole product under cold running
10 s
deionized water.
DISINFECT
ASP CIDEX
within its useful and shelf life, and in compliance with
the manufacturer's warnings and instructions for use.
5 min Immerse the product completely into CIDEX
OPA solution at minimum 20°C. Make certain
that all lumens are filled with disinfectant
solution (use a syringe if needed).
The ASP CIDEX
1 min
a total of three rinses : keep the product
(each rinse)
totally immersed, and use a large volume
of fresh water.
Do not reuse the water for rinsing or any other purpose.
Residues of disinfectant may cause serious side effects.
DRY
Use an air pistol (compressed air) to fully dry the lumen
and the whole product until no more residues of water
are present (visible or detectable).
For example: MediClean Dental, Dr. Weigert Hamburg
4
For example: Bandelin, Sonorex 1028K, 35kHz
5
e.g. 2 gallons
6
MANUAL
. Make certain that
5
OPA solution must be used undiluted,
®
OPA disinfectant requires
®
in
4
®
6