Parameter box
Transport function
Data transfer
Data transfer
After the parameter box has been clipped into place, it
exchanges data with the system screen. It will perform as
programmed in »Standby« (see page 62).
A complete list of the transported data is given in the
»Technical Data« on page 160.
Fully automatic procedure
The message
»Data will be transferred as wanted« is displayed.
After about 12 seconds, all parameter box functions
are active. No user intervention is required.
Semi-automatic procedure
The transport menu is displayed, as preconfigured by
the user in »Standby«.
If the monitoring requirements have changed in the
meantime, the user can intervene in the data transfer
and modify settings in the same way as for the
configuration in »Standby«.
Data transfer can of course also be confirmed
unchanged after checking the display.
These settings are confirmed and then transferred by
pressing the rotary control. After about 12 seconds,
all parameter box functions are active.
List intervals
If the user has opted to transfer list data in the semi-
automatic procedure, the intervals can now be selected.
An interval (monitoring interval) is considered to be the
time between connection and disconnection of the
parameter box.
Each of these intervals are displayed on a separate line in
the list screen, with the identifier »interval« and the date
and time. In addition, the user-programmed device
location is also displayed (see page 64).
88
The patient is new
to the:
Parameter-
box
yes
no
Which data do you want
to transfer from the
parameter box?
Settings
yes
no
Alarm limits
yes
no
List
yes
no
List from interval:
Date
Time
every session
1 18/08
8.00/9.00
2 18/08
13.15/14.00