Dräger Cicero EM Instructions For Use Manual page 74

Integrated anaesthetic workstation
Table of Contents

Advertisement

System screen functions
Softkeys
Ventilation alarm AutoSet
Alarm Info
Ventilation alarm AutoSet
Prerequisite:
The ventilator must be in »IPPV«, »SIMV« or »PCV« mode.
The ventilation alarms can be automatically adjusted to
the current settings of the ventilator by pressing the
»auto set vent. al.« softkey.
The alarm limits previously set in the »limits« menu are
overwritten and cannot be reactivated!
The default alarm limits are not affected by Autoset and
can be reactivated at any time via the functions »config.«,
»alarms« and »default«.
Deactivated alarm limits (– –) are not activated by auto
set!
Parameter new measured value tolerance
upper
P
PEEP + (P
- PEEP) / 4
aw
max
AMV
V
* f * 0.6
T
but always: over 0.5
Alarm Info
The warning and caution fields each have space for
displaying three entries. The advisory message field has
space for four entries.
If further messages are present, a complete list of all
entries in order of priority can be displayed on the
system screen by pressing the softkey »alarm info«.
The list can be read as long as the key is pressed.
74
IPPV alarm limits
standard screen 1- adults
06-12-98
Unit
lower
P
+ 10
mbar
max
V
* f * 1.4 L/min
T
IPPV alarm limits
standard screen 1- adults
06-12-98
Warning:
11:59
The alarm limits of ventilation parameters have been automatically adapted
00:00
to the actual ventilator settings !
Warning:
xxxx
xxxy
xxxz
11:59
00:00
Advisory:
Caution:
limits
auto set
pat. alarm.
auto set
vent. al.
alarm
info
screen
config.
para-
meter
timer
start
config.
Advisory:
Caution:
xxxx
xxxw
xxxy
xxxx
xxxy
xxxz
xxxz
limits
xxxx
xxxy
auto set
xxxz
pat. alarm.
xxxx
auto set
xxxy
vent. al.
xxxz
alarm
info
zzzz
zzzx
screen
zzzy
config.
zzzj
para-
zzzq
meter
timer
start
config.

Advertisement

Table of Contents
loading

Table of Contents