Check operational readiness
Replace cooling air filter of system screen
1 Draw both cooling air filter out of its mount.
Fit new filter or wash in warm water with detergent
and dry thoroughly.
Replace cooling air filter in mount without creasing.
Replace water separator
when contaminated or if message
CO2 LINE ? !
appears on system screen although sample line is not
kinked and does not appear to be blocked. The typical
service life of a water separator is between 2 and
4 weeks.
2 Hold the water separator by its sides and lift out.
Slide the new water separator onto the guide as far as
possible.
Dispose of the old water separator as household
refuse.
Note
Do not dry and reuse the filter. The gas permeability
of the filter is impaired by secretions, nebulisers etc.,
thereby reducing the service life of the filter.