Functional O 2 Saturation (Spo 2 ); Spo 2 Display; Adjusting Settings; Setting Functions - Dräger Cicero EM Instructions For Use Manual

Integrated anaesthetic workstation
Table of Contents

Advertisement

Functional oxygen saturation (SpO
Graph
1 Presentation of the plethysmogram as a curve.
2 Numerical presentation of the saturation value and
pulse rate.
Changing settings
Press the softkey »parameter« on the system screen.
Move the cursor frame to »SpO
rotary control and press to confirm.
The menu for setting the parameters is displayed.

Setting functions

measuring
function
on / off
Activates and deactivates the measuring
function.
pulse
on / off
Activates and deactivates indication of the
pulse rate.
C-Lock
on / off
Activates and deactivates automatic C-Lock
synchronization.
C-Lock synchronizes measurement of the
saturation with the ECG. Leads to better
results if the patient moves or perfusion is
poor.
C-Lock »on« automatically suppresses the
pulse deficit alarm.
mode
slow / normal / fast
changes the speed of measurement.
slow mode: The measured value responds
slowly to changes in oxygen saturation.
Patient movement has little or no effect.
normal mode: The mode for normal
conditions with relatively quiet patients.
fast mode: This mode should be used
whenever rapid reactions are required and
patient movement is negligable.

Coupling with non-invasive pressure measurement

The interlock function should be activated for simulta-
neous measurement of SpO
pressure on one arm. This prevents unnecessary alarms
due to failure to detect a pulse during NiBP measurement
(see page 171).
)
2
/ Pleth.« with the
2
and non-invasive blood
2
Functional oxygen saturation (SpO
Warning:
Man./Spont. alarm limits
standard screen 3 - adults
II
1mV
iBP1
160
0
Pleth
1
123/ 88
06-12-98
11:59
00:00
SpO
%
2
98
Pulse 65
SpO
2
function
on
off
pulse
on
off
C-Lock
(ECG
synchr.)
on
off
mode
slow normal
fast
Parameter box
Caution:
Advisory:
ª
HR/Pulse
ECG
65
1/min
auto set
pulse def 4
pat.alarm
IBP1
ART
mmHg
auto set
122 / 89
vent. al.
alarm
pulse 65
(100)
SpO
%
2
screen
99
config.
pulse 65
meter
2
config.
NiBP
mmHg
°C
T
T
Sys
Dia
1
2
36.8
37.2
(102)
15 min
)
2
limits
info
para-
timer
start
97