Function Keys And Displays - Dräger Cicero EM Instructions For Use Manual

Integrated anaesthetic workstation
Table of Contents

Advertisement

Parameter box

Function keys and displays

Parameter box
Function keys
The parameter box is controlled by the Cicero EM.
A number of functions can be accessed quickly and
G
directly via the function keys.
On pressing this key on the screen or parameter
box, the yellow LED lights up.
The acoustic alarm is then muted for 2 minutes for
Caution and Advisory messages.
¡
Please see the explanations on page 79.
Starts automatic NiBP measurement.
The yellow LED in the key lights up as long as
»Auto« mode remains active.
NiBP measurement can be started and ended
manually with this key.
This key can also be operated between two
measurements in »auto« mode. The time intervals of
automatic measurement are not affected. If the key
is pressed while a measurement is in progress, this
measurement will be interrupted and the cuff
t
deflated.
A static pressure is built up in the NiBP cuff when
this key is pressed in order to stop the venous blood
flow. The yellow LED in the key lights up during this
time.
The blood flow is released when the key is pressed
again. It is released automatically after two minutes.
The pressure is set by the user in the NiBP menu.
Zero alignment key for invasive blood pressure
measurement. The infusion system must be exposed
to atmospheric pressure and the key pressed twice.
If the message »CAL« appears on the screen:
Check the position of the shutoff valves.
Does the pressure transducer still detect an arterial
pulse?
Check pressure transducer.
Displays
1 Act.
Power indicator.
Lights up when the parameter box is active. Shortly
after inserting the parameter box into the holder, the
yellow LED begins to flash until communication with
the system screen has been established.
If the system screen is in Standby the LED does
not light.
86
g
Start
AUTO
Punct.
Stop
ECG
NIBP
SpO
T1
2
d
1
Act.
0
0
T2
P1
P2