Dräger Cicero EM Instructions For Use Manual page 120

Integrated anaesthetic workstation
Table of Contents

Advertisement

Care
Dismantling parts
Breathing system:
Remove flow sensor -
Unscrew pressure limitation valve APL -
Dismantle breathing valves -
valve discs must be handled with care.
Replace measured gas return hose if contamination is
possible -
1 Release the five quick-release screws by turning
through a quarter-turn with a 6 mm hexagon key -
Remove upper part -
2 The two valves are only unscrewed if soiled.
Empty soda lime container (see pages 41 and 128) -
Dispose of with household refuse (not when treating
infectious patients!) -
Piston pump:
3 Release the two quick-release screws by turning
through a quarter-turn with a 6 mm hexagon key -
4 Remove cylinder end cap -
5 Completely unscrew the central screw on the piston
(not quick-release) -
6 Remove piston and roller diaphragm -
The remainder of the piston pump does not come in
contact with breathing gas and does not require
sterilization!
120
1
3
3
4
2
5
6
6