Ambient Conditions; Medical Gas Supply; Vacuum Pressure; Breathing System - Dräger Cicero EM Instructions For Use Manual

Integrated anaesthetic workstation
Table of Contents

Advertisement

Ambient conditions

Temperature .............................: + 15 to + 35 °C
Humidity ...................................: 20 to 90 %, no condensation (operation)
Air pressure..............................: 80 to 106 kPa
CO
content of ambient air
2
in room .....................................: 300 to 800 ppm

Medical gas supply

Pneumatic connections.............: Piped medical gas supply: O
Required pressures
(piped supply)...........................: O
Maximum gas consumption .......: O
O
shortage signal....................: Whistling signal tone
2

Vacuum pressure

Generating principle .................: Ejector system
Power.......................................: Supply (ZV) 3.5 ±0.5 bar: vacuum –0.1...–0.95 bar
Piped vacuum supply (ZV) ........: Vaccum –0.1...–0.95 bar.

Breathing system

Gas volume ..............................: Approx. 3 L total enclosed gas volume
Compliance ..............................: Approx. 3 mL/mbar (approx. 4 mL/mbar with absorber)
Absorber volume ......................: 1.5 L soda lime Drägersorb 800 plus
– 20 to + 50 °C
0 to 98 %, no condensation
50 to 106 kPa
Spare gas cylinders 2 x 3 litres (O
Anaesthetic gas scavenging.
Connection to central vacuum supply system (optional)
: 2.7 to 5.5 bar
2
N
O : 2.7 to 5.5 bar
2
AIR : 2.7 to 5.5 bar
: 100 L/min at 5 bar ± 10 %, including 50 L/min at 5 bar ± 10% for O
2
O : 20 L/min at 5 bar ± 10 %
N
2
AIR : 50 L/min at 5 bar ± 10 %
Triggered: at latest when O
Stops:
at latest when O
Supply (ZV) 5 ±0.5 bar:
Delivery rate approx. 20 L/min (free air intake)
Values dependent on vacuum supplied by piped system (ZV)
Binds approx. 150 L CO
, sufficient for at least 6 hours of operation.
2
(operation)
(storage / transport)
(storage / transport)
(operation)
(storage / transport)
(operation)
, N
O, AIR.
2
2
, AIR) optional
2
pressure drops below 170 kPa (1.7 bar)
2
pressure rises above 260 kPa (2.6 bar)
2
delivery rate 8 L/min (free air intake)
vacuum –0.1...–0.95 bar
delivery rate 13 L/min (free air intake)
Technical data
flush
2
153