Dräger Cicero EM Instructions For Use Manual page 127

Integrated anaesthetic workstation
Table of Contents

Advertisement

Fit the piston pump
Swing up table top
1 Handle in starting position -
Do not damage gaskets when fitting pump!
Hold piston pump by its handle and insert in machine.
2 Swing handle as far as possible to the left (black).
Fit the breathing system
3 Hold breathing system by its handle and insert in
machine -
4 Lock in direction of arrow -
swing as far as possible (black).
Do not damage gaskets when fitting breathing
system!
Fit valves
5 Fit valve discs in expiration and inspiration valves -
fit sealing rings - screw on valve caps.
6 Screw on pressure limitation valve APL -
scale facing the front.
Assemble and fit the inspiratory O
(if used)
Do not touch wire mesh on cell!
Insert cell with conductors facing wiring -
Screw sensor housing together -
7 Plug sensor onto inspiration valve.
Replace O
sensor for gas sampling system
2
(Only if measuring cell is exhausted; display
»FI O
. INOP !«)
2
8 Unscrew the thumbscrew on the right-hand panel of
the device by turning counter-clockwise -
Draw out old sensor and insert new sensor in mount -
Screw thumb screw back into place. The sensor is
automatically calibrated.
sensor
2
Reassembling machine
Reassembling the parts
1
2
3
4
7
6
5
S
G
Q
s
8
u
E
127