System screen functions
Configuring the system screen in »Standby«
Examples of the selectable numerical modules:
HR/Pulse module
Shows the heart rate/pulse rate, which
can be obtained from the ECG, SpO
iBP. The data source is shown in
descending order of priority.
Order of priority:
1. ECG
2. SpO
2
3. iBP 1
4. iBP 2
This module also displays the pulse
deficit (calculated from the difference
between the heart rate and the pulse rate
obtained from SpO
the ST segment height of the 1st ECG.
iBP 1/2 module Shows the systolic/diastolic and mean
values of the invasive blood pressure
measurement. The pulse rate can be
displayed at the same time.
For central venous and intracranial
pressure, the mean value is displayed in
large figures.
CO
module
The end-expiratory CO
2
of the breathing gas (etCO
ted. The inspiratory value (FiCO
also be shown.
Gases module Shows the composition of the breathing
gas in relative concentrations of O
N
O and anaesthetic agent.
2
O
module
Shows the inspiratory O
2
of the breathing gas (FiO
difference in relation to the expiratory
O
concentration (∆ O
2
Comb. haem. module
Shows the combination of measured
values obtained from the measurements
of HR/Pulse, SpO
measuring site).
56
or
2
measurement) and
2
concentration
2
) is indica-
2
) can
2
,
2
concentration
2
) and the
2
).
2
and iBP (only ART
2
H
HR / Pulse (ECG)
65
1/min
pulse def 4
ST +0.05
iBP1
AORTA mmHg
122 / 89
pulse 65
(100)
et CO
Vol.%
2
33
Fi CO
0
2
Vol.% O
N
O
ISO
2
2
29 70 0.8
Fi
25 68 0.6
Fet
Fi O
Vol.%
2
29
∆ O
4.3
2
HR/Pulse
SpO
ART
2
65
98 120/80
1/min
%
mmHg
iBP 2
CVP
mmHg
4.5
mean