Basic Configuration - Dräger Cicero EM Instructions For Use Manual

Integrated anaesthetic workstation
Table of Contents

Advertisement

System screen functions
Configuring the system screen in »Standby«
Record (printer)
This menu is used to configure the interface for data
transfer to a logging printer (see Instructions for Use of
the device to be connected):
record output Determines whether the printer interface
is to be used for a printer or as an
additional MEDIBUS interface.
baud rate
1.2, 2.4, 4.8, 9.6 or 19.2
parity
»even«, »odd« or »none«
data bits
7 or 8
stop bits
1 or 2
If the printer interface and Medibus interface are used
simultaneously, the Baud rate must be set to 1.2.

Basic configuration

The basic configuration comprises three items:
– time
for the current time
– date
for the current date
– language
for the language version.
The same language version is also
used by the ventilator. The following
languages are available:
English
French
German
Dutch
Spanish
Italian
Danish
Swedish
Norwegian
– location
The location of the device can be
entered. If so, it will then be
displayed in the list screen after the
data has been transported with the
parameter box.
Text input for the location:
Select »monitor location«. A frame appears in the
input field.
Letters are displayed inside this frame in alphabetical
order when you turn the rotary control.
Confirm the desired letters. The frame jumps to the
next position where a letter can be entered.
Select the arrow symbol to exit this menu.
64
Standby / Configuration
06-12-98
Standby / Configuration
GB
F
D
06-12-98
NL
E
Language table
I
DK
Language
S
English
N
French
German
Dutch
Spanish
Italian
Danish
Swedish
Norwegian
Settings
mode
alarm limits
parameter
screen
record output
printer
Medibus
acoustic
list entry
baud rate
(kBaud)
1.2 2.4 4.8 9.6 19.2
transport
parity
even
odd
none
RS 232 (Medibus)
data bits
7
8
record (printer)
basic config.
stop bits
1
2
11:59
Printer/Medibus interface configuration menu. Please confirm !
00:00
Settings
mode
alarm limits
parameter
screen
acoustic
list entry
transport
RS 232 (Medibus)
time:
11 : 59
record (printer)
date:
01 - 04 - 96
language:
GB F D NL I E
basic config.
mon. location
11:59
Language, date and time configuration menu. Please confirm !
00:00
Ventilator SW
All
All
All
All
All
All
as from 7.01
as from 7.01
as from 7.20
alarms inactive !
adult
neon.
alarms inactive !
adult
neon.
Monitor SW
All
All
All
All
All
All
as from 4.0
as from 4.0
as from 4.0

Advertisement

Table of Contents
loading

Table of Contents