Message, cause and remedy
Messages on the system screen
(In alphabetical order according to priority)
Messages

Caution messages

AMV
<
!!
>
AMV
!!
<
DIA NIBP
!!
>
DIA NIBP
!!
<
DIAS IBP1
!!
<
DIAS IBP2
!!
>
DIAS IBP1
!!
DIAS IBP2
>
!!
>
ET CO2
!!
ET CO2
<
!!
EXSP-V INOP !!
Fl O2
>
!!
<
Fl HAL
!!
<
Fl ISO
!!
<
Fl ENF
!!
<
Fl DES
!!
<
Fl SEV
!!
FRESH GAS ? !!
>
HEART RATE
!!
>
INSP CO2
!!
104
Cause
Minute volume below lower alarm limit.
Tube blocked / kinked?
Leak in breathing system?
Pressure limited ventilation?
Minute volume exceeds upper alarm limit.
Measured non-invasive blood pressure below
lower alarm limit.
Measured non-invasive blood pressure exceeds
upper alarm limit.
Measured invasive diastolic pressure of the
channel shown (1 or 2) below lower alarm limit.
Measured invasive diastolic pressure of the
channel shown (1 or 2) exceeds upper alarm limit.
Upper alarm limit for end-expiratory CO
concentration has been exceeded for at least two
breaths.
End-expiratory CO
concentration below lower
2
alarm limit for at least two breaths.
Fault in expiration valve. Measured value for minute
ventilation may be excessive.
Inspiratory O
concentration exceeds upper alarm
2
limit.
Applicable inspiratory anaesthetic concentration
below lower alarm limit.
Value below lower alarm limit for at least two
breaths.
Too little gas in manual ventilation bag.
Heart rate exceeds set upper limit.
The inspiratory CO
concentration exceeds the
2
upper alarm limit.
Remedy
Check tube and hoses.
Seal breathing system.
Adjust ventilation pattern.
Correct tidal volume or
respiration rate on ventilator.
Check patient's condition!
Correct alarm limit if necessary.
Check patient's condition!
Correct alarm limit if necessary.
Check patient's condition!
Correct alarm limit if necessary.
Check patient's condition!
Correct alarm limit if necessary.
Check ventilation
2
Check ventilation.
Check for leak in sample line.
Check expiration valve.
Carry out another leak test.
Check O
concentration in fresh
2
gas flow.
This message is deactivated
until the next leakage test by
pressing the alarm suppressor
key.
Check setting of anaesthetic
vaporizer.
Increase fresh gas flow. Repair
any leaks in hose system.
Check patient's condition!
Correct alarm limit if necessary.
Check the soda lime.
Check the valve plate in the
breathing system.
Correct alarm limit if necessary.