Invasive Blood Pressure (Ibp); Start Monitoring; Calibrating Transducers - Dräger Cicero EM Instructions For Use Manual

Integrated anaesthetic workstation
Table of Contents

Advertisement

Parameter box

Invasive blood pressure (iBP)

Invasive blood pressure (iBP)
Start monitoring the invasive blood pressure
The invasive blood pressure can be measured on two
channels simultaneously. The settings and functions of
the channels are identical and are described only once
below.
Connect pressure transducer to the parameter box
(colour-coded grey).
Connect infusion system and catheter/canula to the
transducer. Eliminate any air bubbles in the catheter/
transducer system, as they can impair measurement.
Apply the pressure transducer level with the heart.
If necessary, wait until the pressure transducer has
warmed up.
Connect the transducer dome to the catheter/infusion
system and open it to atmospheric pressure.
Calibrating the sensors
On the system screen
Press the »parameter« softkey.
Select »iBP 1« or »iBP 2« with the rotary control and
press to confirm.
Position the cursor frame over »Calibration« with the
rotary knob and press to confirm. The system
enquires whether the infusion system has been
exposed to atmospheric pressure.
Open the sensor dome and press to confirm.
u
The clock symbol
appears during the calibration
process. A tick ( ) then appears behind the function.
Close the transducer dome and reopen the connec-
tion to the catheter/infusion system; the pressure
trace and pressure values are displayed.
Or on the parameter box:
Press the zero alignment key
ponding sensor twice; a tick ( ) appears behind the
menu item on the system screen.
A question mark (?) appears after the menu item if
a fault occurs during zero alignment. »CAL ?« is
displayed instead of the measured values and the
LED in the
key flashes.
94
of the corres-
g
Start
AUTO
Stop
Warning:
IPPV alarm limits
standard screen 1- adults
2. ECG
II
1mV
iBP1
160
0
CO
40
2
30
0
PAW
20
0
MV
L/min
O
N
O
ISO
2
2
6.0
29 70 0.8
Fi
26 68 0.6
Fet
freq 10
V
0.60
T
06-12-98
11:59
Select a menu position. Please confirm !
Change the settings. Please confirm again !
00:00
g
Start
AUTO
Stop
d
Act.
Punct.
0
0
Advisory:
Caution:
HR/Pulse
(ECG)
ª
65
1/min
iBP 1
function
on
off
pulse def. 4
ST + 0.05
pat. alarm.
iBP1
ART
mmHg
location
AORTA
122/ 89
active
alarms
sys
dia
pulse 65
(100)
etCO
mmHg
graph
absol. morph.
2
33
amplitude
160
Fi CO
0
pulse
2
on
off
Peak
mbar
PEEP
u
calibrate
32
3
plat 26
mean 21
°C
SpO
%
T
T
2
1
2
99
36.8
37.2
pulse 65
d
Act.
Punct.
0
0
limits
auto set
auto set
vent. al.
alarm
info
screen
config.
para-
meter
timer
start
config.

Advertisement

Table of Contents
loading

Table of Contents