Transport Function - Dräger Cicero EM Instructions For Use Manual

Integrated anaesthetic workstation
Table of Contents

Advertisement

System screen functions
Configuring the system screen in »Standby«

Transport function

The parameter box stores the following data, which can
be transferred to the system screen:
– Patient data from the list screen,
– Alarm limit settings and
– Measurement parameter settings.
The patient remains connected to the parameter box
while in transit. The patient's data remains constantly
available. There is no need to reconnect the patient and
set system screens.
The transport function can be preset in »Standby«
(operation in normal use is described on page 83 in the
»Parameter box« section).
Transport function on/off
When switched »off«, no data is exchanged between
the parameter box and the system screen (suitable for
stationary use of the parameter box).
When switched »on«, data is exchanged.
Transport menus yes/no
If switched to »no«, data is transferred between the
parameter box and the system screen in accordance
with the presetting in »Standby«.
If switched to »yes«, the user can modify the type and
scope of data transfer during operation (see also
»Parameter box« section on page 86).
The patient is connected for the first time to ...
If a patient is connected for the first time, the parameter
box will not yet contain any stored data for the patient.
This will generally be the case in the admissions room.
If so, enter »yes« next to »parameter box«.
In response, any remaining data from the previous
patient will be deleted!
If the patient arrives with parameter box already
connected, i.e. with a "previous history" on the device,
enter »no« after »Parameter box«. Data will then be
transferred from the parameter box to the system screen.
Such cases will generally occur in the operating theatre
and in the recovery room.
62
Standby / Configuration
transport
function
on
transport-
menus
yes
The patient is new
to the:
monitor
yes
parameter
box
yes
Which data do you want
to transfer from the
parameter box?
settings
yes
alarm limits
yes
list
yes
Select a menu position. Please confirm !
06-12-98
11:59
Change the settings. Please confirm again !
00:00
alarms inactive !
Settings
mode
adult
neon.
off
alarm limits
parameter
no
screen
acoustic
no
list entry
transport
no
RS 232 (Medibus)
record (printer)
basic config.
no
no
no

Advertisement

Table of Contents
loading

Table of Contents