Transport function

The parameter box stores:
– the patient data from the list screen
– alarm limit settings and
– measurement parameter settings.
The patient remains connected to the parameter box in
transit. The patient's data remains constantly accessible.
There is no need to reconnect the patient and configure
the monitors.
Plugging in/unplugging the parameter box
White housing:
To connect:
Hold the parameter box by the blue handle.
Slide the parameter box into the white housing.
It clicks audibly into place, and all LEDs on the front
panel light up briefly.
To disconnect:
Pull out the blue handle. After overcoming a tangible
resistance, the box is electrically disconnected and
freely transportable.
Blue holder:
The parameter box can be tilted in the blue holder. On
overcoming a set resistance, it can be adjusted to an
ergonomic position.
To connect:
Hold the parameter box by the blue handle.
Place the parameter box on the crossbar of the holder
and allow it to swivel down.
It clicks audibly into place, and all LEDs on the front
panel light up briefly.
To disconnect:
Pull out the blue handle. Swivel upwards to discon-
nect electrically, and lift off the crossbar. The box is
now freely transportable.
Parameter box
Transport function
Fixing the parameter box to the unit