Integrated anaesthetic workstation
Table of Contents

Advertisement

Messages on the system screen
(In alphabetical order according to priority)
Messages
<
MEAN IBP1
!!
<
MEAN IBP2
!!
>
MEAN IBP1
!!
>
MEAN IBP2
!!
PRESS LIMIT !!, (!)
>
PULSE IBP 1
!!
>
PULSE IBP 2
!!
>
PULS SPO2
!!
>
S-T MV
!!!
SPO2
>
!!
<
SYS IBP1
!!
<
SYS IBP2
!!
>
SYS IBP1
!!
>
SYS IBP2
!!
<
SYS NIBP
!!
>
SYS NIBP
!!
TEMP 1
<
!!
<
TEMP 2
!!
>
TEMP 1
!!
>
TEMP 2
!!
Cause
The average invasively measured blood pressure of
the displayed channel (1 or 2) is below the set
lower alarm limit.
The average invasively measured blood pressure of
the displayed channel (1 or 2) is below the set
lower alarm limit.
Ventilator operating with pressure limitation.
Change in lung compliance? Tube buckled?
Microbial filter in inspiratory line blocked?
Invasive measured pulse exceeds set alarm limit.
Pulse rate exceeds upper alarm limit.
Magnitude of S-T segment exceeds upper alarm
limit.
Oxygen saturation exceeds set upper alarm limit.
Measured invasive systolic pressure of the channel
shown (1 or 2) below set alarm limit.
Measured invasive systolic pressure of the channel
shown (1 or 2) exceeds set alarm limit.
Measured non-invasive systolic pressure below set
alarm limit.
Measured non-invasive systolic pressure exceeds
set alarm limit.
Measured body temperature of the channel shown
(1 or 2) below set alarm limit.
Measured body temperature of the channel shown
(1 or 2) exceeds set alarm limit.
Message, cause and remedy
Remedy
Check patient's condition!
Correct alarm limit if necessary.
Check patient's condition!
Correct alarm limit if necessary.
Check tube/filter.
Increase Pmax or decrease V
necessary.
If pressure-limited ventilation is
intentional: press the »config«,
softkey and select »alarms« ; in
the alarms menu, switch »Pmax«
to »off« , to reduce the warning
message to a caution message
with no audible alarm.
Check patient's condition!
Correct alarm limit if necessary.
Check patient's condition!
Correct alarm limit if necessary.
Check patient's condition!
Correct alarm limit if necessary.
Check O
concentration in fresh
2
gas.
Check ventilator.
Check patient's condition!
Correct alarm limit if necessary.
Check patient's condition!
Correct alarm limit if necessary.
Check patient's condition!
Correct alarm limit if necessary.
Check patient's condition!
Correct alarm limit if necessary.
Check patient's condition!
Correct alarm limit if necessary.
Check patient's condition!
Correct alarm limit if necessary.
if
T
105