Ventilator; Gas Delivery - Dräger Cicero EM Instructions For Use Manual

Integrated anaesthetic workstation
Table of Contents

Advertisement

Technical data
Pressure relief ..........................: 90 ± 10 mbar
Pressure limiting valve...............: Variable between 5 and 70 mbar, ± 15% of set value
Patient connections ..................: ISO cone 22 mm
Anaesthetic gas scavenging......: Nominal socket diameter 27 mm or ISO cone 30 mm
Leakage ventilation system........: < 120 mL/min at 30 mbar
Maximum pressure in
breathing system ......................: 10 to 80 mbar ± 20 % or + 3 mbar, but at least 10 mbar over PEEP.
Expiratory resistance.................: 4 mbar at a flow of 60 L/min (conforming to EN 740).
Inspiratory resistance................: 4 mbar at a flow of 60 L/min (conforming to EN 740).
Non-return valves......................: Back-pressure of a wet valve < 1 mbar

Ventilator

Drive gas consumption .............: 2 L/min O
Tidal volume (V
).......................: <20 mL
T
Frequency (f
).......................: 6 to 80 per minute (as from software version 7.2)
IPPV
*
Minute ventilation (
)..............: Up to 25 L/min
E
T
:T
.........................................: 1:3 to 2:1
I
E
T
:T
........................................: 0 to 60 %
IP
I
*
Inspiratory flow
max ..............: Max. 75 L/min
I
...........................................: 3 to 80 per minute ± 5 % (f
f
IMV
PEEP........................................: 0; 2 to 15 mbar
Trigger pressure .......................: – 1 mbar
Inspiration flow in PCV mode ....: 5 to 75 l/min (PCV ventilator)
Operating pressure in PCV mode : 10 to 70 mbar (PCV ventilator)

Gas delivery

Low-flow measuring tubes ........: (Calibrated for 20 °C, 1013 mbar)
Delivery output..........................: O
154
;
Opening pressure of a wet non-return valve < 1 mbar
or AIR in operation
2
0 L/min in »Standby« mode
± 30 % or ± 6 mL (as from software version 7.2)
± 10 % or ± 10 mL
20 to 100 mL
± 5 % or ± 15 mL
100 to 1400 mL
6 to 60 per minute (software version < 7.2)
± 5 %
± 5 %
3 to 60 per minute ± 5 % (f
± 10 % or ± 2 mbar
(P
max
± 10 % or ± 2 mbar
0; 2 to 15 mbar
(P
max
± 0,5 mbar (invariable)
O
:
0.02 to 0.5 L/min
2
At minimum scale:
At minimum scale:
O
:
0.55 to 10.0 L/min
2
At minimum scale:
N
O:
0.02 to 0.5 L/min
2
At minimum scale:
At minimum scale:
N
O:
0.55 to 10.0 L/min
2
At minimum scale:
AIR:
0.2 to 14.0 L/min
At minimum scale:
, N
O, AIR at least 9 L/min each
2
2
< f
) (as from software version 7.2)
IMV
IPPV
< f
) (software version 7.2)
IMV
IPPV
- PEEP > 5 mbar, as from software version 7.2)
- PEEP > 10 mbar, software version 7.2)
± 10 %
+ 20 %, – 10 %
± 5 %
± 10 %
+ 20 %, – 5 %
± 10 %
– 20 %, + 10 %
± 5 %
± 10 %
+ 20 %, – 5 %
± 10 %
+ 15 %, – 5 %
;
=
(
EN 1281-1)