Dräger Cicero EM Instructions For Use Manual page 188

Integrated anaesthetic workstation
Table of Contents

Advertisement

Description
Operation of the ceiling version
1 Move Cicero EM away from the DVE.
2 Swing supporting flap upwards.
The DVE can now be controlled via its own keypad
again.
As from approx. 5 cm above the lower limit position,
the DVE can also be moved upwards even when the
supporting flap is lowered.
What to do if problems arise:
If the machine inadvertently strikes an obstacle before
reaching the floor, the DVE will stop before it is
forcibly disconnected and cannot be controlled via
the keypad on the Cicero EM.
The machine can then only be raised via the keypad
on the DVE.
If the machine subsequently remains at an angle on
its mounting journal:
Briefly lower it onto the ground and pick it up again
with the DVE.
The mount may suffer damage in extreme cases!
Call DrägerService if damage is suspected!
Depending on how serious the damage is, the Cicero
may crash to the ground when raised again!
Maintenance and care
The windows on the optical sensors must be cleaned!
They are resistant to all cleaning agents and
disinfectants usually used in hospitals.
Abrasive agents must not be used.
188