Dismantling parts

Measured gas:
1 T-piece of measured gas hose (if used): remove filter.
Inspiratory O
sensor (if used):
2
Must not be disinfected in liquid or autoclaved!
Wipe any dirt off housing or lead with damp
disposable cloth.
Unscrew O
sensor and remove O
2
touch wire mesh of O
2
water-repellent property and measurement becomes
inaccurate.
Dirt on wire mesh of sensor cell is simply wiped off
with a disposable cloth moderately soaked in distilled
water.
Temperature sensor: (breathing gas; if used)
Draw temperature sensor out of Y-piece, remove
Y-piece from ventilation hoses.
Prise lead out of hose clips.
Pull connector out of socket.
Wipe dirt off with damp disposable cloth.
Flow sensor
Contamination by infectious patients: dispose of as
directed!
If there is no risk of contamination, continue using the
flow sensor until the message »FLOW INOP !« appears at
the breathing minute volume. Then dispose of the flow
sensor as ordinary domestic waste.
It can be incinerated at temperatures above 800 °C with
low pollutant emissions.
cell. Do not
2
cell, otherwise it loses its
1
Care
Dismantling parts