repeated 5 seconds later with a
higher cuff pressure. If it is still too
low, the air is released again and the
measurement repeated 30 seconds
later with an elevated cuff pressure.
Punction pressure numerical display
(Separate for adults and neonates).
Sets the punction inflation pressure
in the sphygmomanometer cuff.
Turn the rotary knob to change the
value, and press to confirm.
Unit mmHg / kPa Switching over between the
measuring units »mmHg« and »kPa«.
Interlock
NiBP–SpO
on / off
2
Activates and deactivates the inter-
lock function between »NiBP« and
»SpO
2
When the NiBP sphygmomanometer
cuff is inflated with the Interlock on,
the »SpO
deactivated in order to prevent false
alarms if both measurements are
taken on the same arm. Generally,
»Interlock on« is recommendated.
Interlock
NiBP–iBP
auto / off
Activates and deactivates the
interlock function
between »NiBP« and »iBP«.
The »iBP 1/2 PULSE ? !!!«
measurement is suppressed while the
cuff is inflated.
NiBP alarms:
NiBP alarms are suppressed during invasive blood
pressure measurement in an artery.
G
If an NiBP alarm is triggered, the alarm sound is
suppressed until the next NiBP measurement by
pressing
.

Measuring the blood pressure of neonates

The parameter box includes special algorithms in
neonatal mode for monitoring neonates. Changing over
to neonatal mode is described on page 49.
«.
PULSE ? !!!« alarm is
2
Non-invasive blood pressure (NiBP)
NiBP
auto
on
off
punction
Start Stop
measure
Start Stop Turbo
time interval
5
active alarms
sys dia
initial pressure
180 mmHg
punction pressure 40
mmHg
units
mmHg kPa
interlock:
NiBP-SpO
on
off
2
NiBP-iBP
on
off
Standby / Configuration
06-12-98
11:59
Select the required ventilation mode (adult or neonate). Please confirm !
00:00
Parameter box
alarms inactive !
Default values
mode
adult
neon.