Parameter box
Non-invasive blood pressure (NiBP)
Changing settings
Press the softkey »parameter«.
Move the cursor frame to »NiBP« with the rotary
control and press to confirm.
The menu for setting the parameter is displayed.
auto
on / off
Activates automatic measurements at
the specified time intervals.
punction
Start / Stop
Starts or stops venous congestion.
Inflation ends automatically after
2 minutes.
measurement
Start / Stop / Turbo
»Start« Triggers manual measure-
»Stop« Stops a measurement.
»Turbo« Five simplified measure-
time interval
2 / 3 / 5 / 10 / 15
Selects the time interval for automatic
measurement in minutes. The starting
time for the next automatic measure-
ment is not affected by a measure-
ment triggered manually between two
automatic measurements.
active alarms
sys / dia
Determines whether the systolic or
diastolic pressure is to be monitored.
punction pressure numerical
(separate for adults and neonates)
Sets the punction inflation pressure
of the sphygmomanometer cuff.
The cuff is inflated to the preset
starting pressure for the first
measurement; after reading off the
systolic pressure, the cuff is inflated
to 20 mmHg above the last
measured systolic value for the next
measurement. If the inflation
pressure is too low for correct
measurement of the systolic
pressure, air is automatically released
from the cuff and the measurement is
92
ment outside the automatic
time interval.
Or press the
key on
the parameter box.
ments in five minutes.
This mode can also be
stopped with »Stop«.
Warning:
IPPV alarm limits
standard screen 1- adults
2. ECG
II
1mV
iBP1
160
0
CO
40
2
30
0
PAW
20
0
MV
L/min
O
N
O
ISO
2
2
6.0
29 70 0.8
Fi
26 68 0.6
Fet
freq 10
V
0.60
T
06-12-98
11:59
Select a menu position. Please confirm !
00:00
Change the settings. Please confirm again !
NiBP
auto
on
off
punction
Start Stop
measure
Start Stop Turbo
time interval
5
active alarms
sys dia
initial pressure
180 mmHg
punction pressure 40
mmHg
units
mmHg kPa
interlock:
NiBP-SpO
on
off
2
NiBP-iBP
on
off
Advisory:
Caution:
HR/Pulse
(ECG)
ª
65
1/min
NiBP
auto set
auto
on
off
pulse def. 4
ST + 0.05
pat. alarm.
punction
Start Stop
iBP1
ART
mmHg
measure
Start Stop Turbo
122/ 89
time interval
5
pulse 65
(100)
active alarms
sys dia
etCO
mmHg
2
initial pressure
180 mmHg
33
punction pressure 40
mmHg
Fi CO
0
2
units
mmHg kPa
Peak
mbar
PEEP
interlock:
32
3
NiBP-SpO
on
off
2
NiBP-iBP
on
off
plat 26
mean 21
°C
SpO
%
T
T
2
1
2
99
36.8
37.2
pulse 65
limits
auto set
vent. al.
alarm
info
screen
config.
para-
meter
timer
start
config.