Check Vapor; Check Anaesthetic Gas Scavenging - Dräger Cicero EM Instructions For Use Manual

Integrated anaesthetic workstation
Table of Contents

Advertisement

Preparing for use
Checking workstation with checklist

Check Vapor:

– Set the handwheel to »
– Filling level OK?
– Last inspection less than six months ago?
Safety filler:
– The plug is inserted and secured.
Vaporiser mount:
– Vaporiser is correctly seated.
Locking mechanism:
– Plug-in system is locked (lever to the left as far as
possible).

Check anaesthetic gas scavenging:

– Is indicator in wall socket green? (Only when using
Dräger systems; note sounds of gas flow in other
cases.)
– Hose connected to breathing system?
– Is the anaesthetic gas scavenging system correctly
installed?
– Are the two hoses fitted correctly?
– Are the vents open?
– Are all hoses free of kinks?
1 Is the transfer hose of the anaesthetic gas scavenging
system free of condensate? Drain if necessary.
22
O
«.
Check-list Cicero EM
To be verified by the user prior to each use
Vapor
handwheel set to zero
fill - level ok
less than 6 months since last inspection
correctly seated and locked
Safety-fill
plug secured in place
0
Vol.
%
Safety f
Gas scavenger
operational
Breathing system
complete, locked in place
Soda lime
no colour change
Emergency breathing bag
present, functional
C t l
l
ll
t
th
D
1
p ug secu ed n p ace
2 7 b