Measuring ranges
for graph display .......................: 20, 40, 80, 160 and 320 mmHg
Max. pressure signal .................: – 10 to 320 mmHg for undistorted display
Measured parameters ...............: Systolic, diastolic and mean pressure
Measuring range.......................: – 10 to 320 mmHg
Resolution ................................: 1 mmHg
Unit of measure ........................: mmHg
Site labels.................................: ART, AORTA, APulm, CVP, ICP, ?
Oxygen saturation (SpO
2
Measurement of functional SpO
Input.........................................: IEC type CF, floating, Colour-coded brown.
Cable type................................: without preamplifier.
Wavelengths.............................: 660 nm (red) and 920 nm (infrared)
Measuring range.......................: 0 to 105 %
Saturation resolution .................: 1%
Accuracy ..................................: Adults:
Accuracy of Durasensor ...........: ± 3 % (additional)
Measuring modes .....................: Normal:
Plethysmogram.........................: Automatic amplitude control.
Pulse rate .................................: Derived from plethysmogram; no pulse rate tone in slow mode. The pulse rate tone
Pulse rate resolution .................: 1/min
Pulse rate accuracy ..................: ± 1/min

Body temperature measurement

Inputs .......................................: Floating, IEC type CF, colour-coded green.
Parameter.................................: Two temperature values (T1, T2); both are monitored by alarm limits.
Measuring ranges .....................: 0 °C to 45 °C;
Resolution ................................: 0.1 °C.
Accuracy ..................................: ±0.1 °C.
Self-test....................................: A fault message is output when one of the automatically tested fault limits is
adjustment of the measuring label, alarm limits and display configuration are
automatically controlled by the site label.
)
saturation.
2
0 to 50 %
50 to 70 %
70 to 100 %
Neonates:
0 to 70 %
70 to 95 %
95 to 100 %
measurement over 5 to 7 seconds
Fast:
2 to 3 seconds
Slow:
10 to 15 seconds
Off:
Module deactivated
is controlled by the SpO
exceeded.
undefined
± 3 %
± 2 %
undefined
± 3 %
not specified
measurement when programmed accordingly.
2
Technical data