Steam Sterilization (121 / 134 °C) - Dräger Cicero EM Instructions For Use Manual

Integrated anaesthetic workstation
Table of Contents

Advertisement

Care
Disinfection / Cleaning / Sterilization
Agents and methods
Steam sterilization at 121 or 134 °C
Cicero parts which can be sterilized in this way are listed
in the table on page 125.
Steam sterilization must be preceded by sterilization with
moist heat and/or wipe cleaning. All parts must be dry!
Steam sterilization is only required after treating an
infectious patient.
Minimum exposure times: 20 minutes at 121 °C and
10 minutes at 134 °C. Plastic and rubber parts are
subject to greater material wear when sterilized at
134 °C.