Message, cause and remedy
Messages in the ventilation system
Message must be acknowledged
Messages during
the self-test
Service No. 118
• Incorrect test control pressure. No sealing
Service No. 119
• Test control pressure too high.
Service No. 120
• Operating control pressure too high.
Service No. 214
• Cold/warm restart recognition not
Standby
Subsystem 1/2 leak
• Leak in subsystem 1 and/or 2.
Subsystem 1 leak
• Leak in area of inspiration valve, valve 7,
Subsystem 2 leak
• Leak in area of inspiration valve, absorber, Close quick-release couplings.
Subsystem 3 leak
• Leak in area of breathing bag,
Test ..x interrupted
Valve plate faulty
• Breathing system valve plate or
Valve plate faulty
• Breathing system valve plate or
Valve plate faulty
• Breathing system valve plate or
Version ....
Y-piece closed ?
• Dialogue during the self-test
Y-piece open ?
• Dialogue during the self-test
????
Last interruption
Cause
test possible for valve and roller diaphragm.
guaranteed.
The machine is ready for operation.
(Leak greater than 5 L/min at 30 mbar)
patient hose, flow sensor or O
and PEEP valve (V
)
6
piston pump, valves 1-2-3, piston
pump/breathing system interface and
PEEP valve (V
)
6
fresh gas hose, Vapor
Displayed after the self-test is stopped.
diaphragm of V
defective.
1
Display »112« in V
field
T
diaphragm of V
defective.
2
Display »113« in V
field
T
diaphragm of V
defective.
3
Display »114« in V
field
T
Display at the start of the self-test
Internal electronic malfunction.
The self-test has already been stopped
9 times!
Remedy
Call DrägerService.
Call DrägerService.
Call DrägerService.
Call DrägerService.
Select operating mode.
Check subsystems.
(see chart on page 164).
Close the quick-release couplings.
sensor
Connect the P-sensor. Replace the valve disc
2
in the inspiration valve.
Check connections and breathing
bag or hoses. Check connector system and
sealing rings on Vapor.
Select operating mode.
Replace breathing system or valve plate.
Pneumatic interface faulty (check lip seals)
Replace breathing system or valve plate.
Pneumatic interface faulty (check lip seals).
Replace breathing system or valve plate.
Pneumatic interface faulty (check lip seals).
Software version No. is displayed after
switching on. In the event of malfunctions,
quote the version number to DrägerService.
Close the Y-piece for the remainder of the
self-test.
Open the Y-piece for the internal pressure
sensor calibration. If using an external
humidifier, remove the breathing hose from
the inspiration cone or filter.
Call DrägerService.
For information. The next time you start up the
machine, the complete self-test will have to
be run.