Message, cause and remedy
Messages on the system screen
(In alphabetical order according to priority)
Messages
PRESSURE INOP !
RS 232 COM !
SELECT AGAS !
G
SPO2
OFF !
SPO2 INOP !
SPO2 SENSOR? !
TEMP1 CALIB !
TEMP2 CALIB !
TEMP 1 INOP !
TEMP 2 INOP !
WATER TRAP ? !
108
Cause
Pressure sensor defective.
RS 232 communication interrupted.
The automatically recognised anaesthetic agent
does not match the manually selected anaesthetic.
SpO
alarm inactive.
2
SpO
measurement faulty.
2
No pulse detected by pulsoxymeter although heart
is clearly beating.
Fault found during automatic internal accuracy test. Call DrägerService
Temperature measuring function defective.
The water trap is full or severely fouled.
No measured values are displayed for CO
anaesthetic agent and O
.
2
Remedy
Call DrägerService
Check plug connection.
Check the anaesthetic agent
used or switch over to automatic
recognition.
Call DrägerService
Check patient's condition
(disturbed circulation?)
Check SpO
sensor positioned
2
correctly.
Replace temperature sensor.
Call DrägerService
Empty the water trap.
, N
O,
Check the fitting of the
2
2
collecting jar. Remove
condensate from the prism.