Notes on Quick Start
The Quick Start procedure is permitted a maximum of
10 times in succession, and only if the previous (full) self-
test was completed without error.
Otherwise, the display on the ventilator and the system
screen instruct the user to run through the complete self-
test.
Manual ventilation remains possible under all condi-
tions!
Quick Start can also be triggered by the »Cancel
Test« softkey.
Quick Start can be triggered at any time, even during
a self-test.
If there is a power failure
possible)
Make sure that the master switch is pressed down
– The acoustic power failure warning is muted after
30 seconds.
Deliver fresh gas - set APL valve
– If required, press the O
Note: If there is a power failure, the piston of the
ventilator is pushed to its end position by the
airway pressure. This increases the system volume
by a maximum of 1.4 litres.
If there is a gas failure
In case of AIR failure
(medical air)
– The system automatically switches over to O
supply.
In case of O
failure
2
– The system automatically switches over to the
AIR supply. An acoustic alarm is sounded (O
shortage signal). The supply of N
If both O
and air fail
2
The patient must immediately be ventilated with the
separate emergency ventilation bag!
(manual ventilation is
+
flush button (»O
«).
2
2
2
2
O is disabled.
2