Patient alarm AutoSet
Alarm limits can also be set automatically once the
patient is stable.
The user can set a new limit range with new upper and
lower alarm limits by pressing the softkey »auto set pat.
al.«.
The alarm limits set in the »limits« menu are deleted and
cannot be automatically reactivated. The default alarm
limits remain unaffected by the Autoset and can be
reactivated at any time via the functions »config.«,
»alarms« and »default«.
The alarm limits for SpO
2
unaffected.
AutoSet cannot set the lower alarm limits for the heart
rate below 40 beats per minute and below 90 mmHg for
systolic pressure values (50 mmHg for neonates).
Deactivated alarm limits (– –) are not activated by
auto set!
Parameter
new measured value
tolerance ...
upper
ECG/Pulse HR/Pulse
+20.0
ST
+0.05
NiBP/ART SYS
+20.0
DIA
+20.0
AORTA
SYS
+20.0
DIA
+20.0
ART pul.
SYS
+15.0
DIA
+5.0
MIT
+10.0
CVP
MIT
+5.0
ICP
MIT
+5.0
mmHg
SYS
+20.0
DIA
+20.0
MIT
+20.0
TEMP
+1.0
etCO
+5.0
2
Fi AGAS
HAL
+0.5
ENF
+0.5
ISO
+0.5
DES
+0.5
SEV
+0.5
and FiO
also remain
2
Unit
lower
–20.0
beats per
minute
–0.05
mV
–20.0
mmHg
–20.0
mmHg
–20.0
mmHg
–20.0
mmHg
–15.0
mmHg
–5.0
mmHg
–10.0
mmHg
–5.0
mmHg
–5.0
mmHg
–20.0
mmHg
–20.0
mmHg
–20.0
mmHg
°C
–1.0
–5.0
mmHg
–0.5
Vol. %
–0.5
Vol. %
–0.5
Vol. %
–0.5
Vol. %
–0.5
Vol. %
Warning:
IPPV alarm limits
standard screen 1- adults
06-12-98
11:59
The alarm limits of haemodynamic parameters have been automatically adapted
00:00
to the actual measurement values !
Remarks
but not less than 40 beats per minute
but not less than 90 mmHg (50 mmHg for neonates)
System screen functions
Softkeys
Patient alarm AutoSet
Advisory:
Caution:
limits
auto set
pat. alarm.
auto set
vent. al.
alarm
info
screen
config.
para-
meter
timer
start
config.
73

Advertisement

Table of Contents
loading

Table of Contents