The MAN/SPONT alarm limits are automatically activated
in »MAN/SPONT« mode and the SIMV alarm limits for
»SIMV« mode with a ventilation frequency of less than
6 breaths per minute. The currently active alarm limits are
indicated in the status field and flash for the first five
seconds.
A special alarm mode can be activated via the function
»alarms« in the »config.« menu for operation with a
heart/lung machine (»HLM« mode). In this case, all apnea
alarms are deactivated and the gas values are indicated
continuously, regardless of the respiratory phase.
All other alarm monitoring continues as in the ventilation
mode from which »HLM« was activated.
The message »HLM« is also displayed on screen after the
ventilation mode.
Alarm limits in ventilation mode
Minute volume
Inspiratory anaesthetic agent
Airway pressure
Inspiratory O
concentration
2
Apnea pressure
Apnea flow
Apnea CO
2
All other alarm limits remain unaffected by this function.
IPPV alarm limits
standard screen 1- adults
2. ECG
II
1mV
iBP1
160
0
CO
40
2
30
0
PAW
20
0
O
29 70 0.8
Fi
26 68 0.6
Fet
06-12-98
»MAN/SPONT«
>
AMV
off/on
<
off/on
>
on
<
off/on
>
Paw
on
<
off
>
FiO
off/on
2
<
on
off
off
set to 1 minute
Warning:
MV
L/min
SpO
N
O
ISO
2
2
99
6.0
pulse 65
freq 10
V
0.60
T
11:59
00:00
»SIMV« with f
< 6
IMV
off/on
off/on
on
off/on
on
on
off/on
on
set to 30 seconds
set to 30 seconds
set to 30 seconds
System screen functions
Alarm concept
Caution:
Advisory:
HR/Pulse
(ECG)
ª
limits
65
1/min
auto set
pulse def. 4
ST + 0.05
pat. alarm.
IBP1
ART
mmHg
auto set
122 / 89
vent. al.
alarm
pulse 65
(100)
info
etCO
mmHg
2
screen
33
config.
para-
meter
Peak
PEEP
timer
32
3
start
plat 26
mean 21
config.
°C
%
T
T
2
1
2
36.8
37.2
36.8
T1
37.5
T2
≥ 6
»SIMV« with f
IMV
for »IPPV« and »PCV«
off/on
off/on
on
off/on
on
on
off/on
on
set to 15 seconds
set to 15 seconds
set to 15 seconds
81