System screen functions
Configuring the system screen in »Standby«
pEEG modules pEEG modules are available for
selection in the parameter menu
(see page 52).
Please observe the separate Instructions
for Use!
Volumeter module
This module comprises a breath volumeter (upper bar
graph) and a minute volumeter (lower bar graph). The
length of the bar represents the maximum measured
value in accordance with the mode selected:
Mode
Neonates
Adults
Breath volumeter:
– The expiratory volume of each breath (V
graphically by the upper bar and numerically in large
numerals abore.
Minute volumeter:
– The minute volumeter in the lower bar graph runs for
60 seconds, during which it adds up the tidal
volumes.
The time expired is shown in seconds together with the
value in large numerals. The individual breaths are shown
as segments in the bar graph. The volumeter stops
automatically after 60 seconds and beeps. The measured
values are displayed for 4 minutes and then deleted.
To start the volumeter:
Press the rotary control. If it is pressed again during
the 60 seconds, the values are deleted and the
volumeter restarted.
58
Breath V
MV
T
0.25 L
2.5 L/min
1.25 L
12.5 L/min
) is shown
T
SEF
Hz
SMF
90
12.5 10.5
EMG
SEF / SMF / EMG
SEF
Hz
d-Ratio
90
12.5 0.5
EMG
SEF / d-Ratio / EMG
L.SEF
Hz R.SEF
90
90
12.5 10.5
EMG
L.SEF / R.SEF / EMG
Total
Ampl.
delta
47 µV 34
BSR
Amplitude
0.60 L
V
T
1.25
4.3 L
30 s
Volumeter
12.5
SMF
Hz
d-Ratio
0.5 10.5
EMG
SMF / d-Ratio / EMG
SEF
Hz
d-Ratio
90
12.5 0.5
d/q/a/b
SEF / d-Ratio / d / t / a / b
L.SEF
Hz R.SEF
90
90
12.5 10.5
L.SMF
R.SMF
9.5
8.5
L.SEF / R.SEF / SMF