Dräger Cicero EM Instructions For Use Manual page 103

Integrated anaesthetic workstation
Table of Contents

Advertisement

Messages on the system screen
(In alphabetical order according to priority)
Messages
>
Fl HAL
!!!
>
Fl ISO
!!!
>
Fl ENF
!!!
>
Fl DES
!!!
Fl SEV
>
!!!
<
Fl O2
!!!
HR RATE
<
!!!
IBP 1 PULSE ? !!!
IBP 2 PULSE ? !!!
PAW
>
!!!
PAW NEGATIVE !!!
EXP PRESSURE
>
!!!
PULSE IBP 1
<
!!!
<
PULSE IBP 2
!!!
<
PULSE SPO2
!!!
SPO2
<
!!!
SPO2 PULSE ? !!!
VENT INOP !!!
Cause
The inspiratory anaesthetic concentration exceeds
the upper alarm limit in each case.
The upper alarm limit has been exceeded for at
least two breaths.
The inspiratory O
concentration is below the
2
lower alarm limit.
The heart rate is below the set limit.
No pulse detected during invasive measurement.
Airway pressure exceeds the upper alarm limit.
Ventilation hose buckled? Stenosis?
Pulmonary problem? Cough?
Negative airway pressure measured.
End-expiratory pressure exceeds PEEP by more
than 10 mbar.
The measured invasive pulse is below the set
alarm limit.
Pulse rate has dropped below the set
alarm limit.
Oxygen saturation is below the set lower
alarm limit.
No pulse signal detected by SpO
for approx. 10 seconds.
Fault in ventilator pressure sensor.
Filter in pressure measuring hose blocked (water
absorption!)
Control pressure in ventilator too low.
Hardware fault in ventilator.
Machine not operational!
Remedy
Check setting of anaesthetic
vaporizer.
Check O
Check setting on O
Check patient's condition!
Correct alarm limit if necessary.
Check patient's condition!
Check patient's condition!
Check hose system, tube and
fresh gas flow.
Check hose system and tube on
ventilator.
Check anaesthetic gas
scavenging line.
Extend the expiration time.
Check the hose system and
microbial filter.
Drain the water traps.
Check patient's condition!
Check patient's condition!
Check ventilation.
Check inspired O
tion.
measurement
Read ECG.
2
Check SpO
off? NiBP measurement on
same arm?)
If measured on the same arm as
NiBP: Switch »on« Interlock!
Immediately ventilate the
patient manually!
If no pressure build-up in the
breathing system: immediately
ventilate the patient with the
separate emergency breathing
bag.
Check medical gas supply.
Replace filter in pressure
measuring hose.
Message, cause and remedy
supply.
2
flow meter.
2
concentra-
2
sensor (dropped
2
103