Dräger Cicero EM Instructions For Use Manual page 6

Integrated anaesthetic workstation
Table of Contents

Advertisement

Emergency Quick-Start
Emergency Quick-Start
Press the master switch (1)
– The pressure gauges (2) are in the green zone.
– All lamps on the ventilator light up.
Press a function key on the ventilator
Recommendation: Press and hold down
longer than 1 second.
– The message »Test interrupted« is displayed.
For devices without PCV mode (software 6.n):
Deliver fresh gas
– With the delivery valve buttons (3) under the
measuring tubes
– If necessary fill the system rapidly with the O
flush (4).
– The manual ventilation bag should fill.
Set the APL pressure limiting valve (5) to »MAN«
Set maximum pressure
– Turn the valve head (6) until the disc is at the
desired maximum pressure.
6
M
key for
2
5
6
D
2
I
O
MAN
SPONT
IPPV
P
max
T
:T
IP
I
SIMV
PCV
V
f
IMV
T
f
IPPV
T
I
:T
E
PEEP
P
PCV
Flow
O
4
AIR
2
3
MAN
SPONT
IPPV
Lecktest ?
P
max
T
IP
:T
I
SIMV
PCV
V
f
T
:T
PEEP
f
IMV
T
IPPV
I
E
P
PCV
Flow
Gas/Power
ON / I
Press
1
TEST
O
TEST