Dräger Cicero EM Instructions For Use Manual page 52

Integrated anaesthetic workstation
Table of Contents

Advertisement

System screen functions
Configuring the system screen in »Standby«
SpO
/Pleth settings
2
(see also page 97):
measuring function
Switches SpO
pulse
Switches pulse rate display on and off.
C-Lock
Switches ECG synchronization on and
(ECG synchr.) off.
mode
Determines the speed of measurement.
Fast measurements are more susceptible
to antefoets.
CO
/O
settings
2
2
sample rate
Selects the flow rate of the sample gas
through the sample gas line at the Y-
piece. Dräger recommends 200 mL/min
with connected sample gas return line.
CO
amplitude The maximum amplitude of the CO
2
time curve can be set here.
CO
unit
The CO
2
2
between mmHg and kPa.
insp. CO
Switches the numerical display of the
2
inspiratory CO
Inspiratory CO
mode with an adjustable alarm limit of
0 to 10 mmHg.
sidestream O
measurement
2
»on«:
O
sidestream.
»off«:
O
O
(requires inspiratory O
pEEG settings
This menu selects the pEEG Parameters displayed in the
pEEG module (see page 58).
52
measurement on and off.
2
2
measuring unit can be selected
concentration on or off.
2
is monitored in IPPV
2
is measured by sampling the
2
is measured by measuring
2
at the inspiratory port
2
-sensor).
2
Standby / Configuration
SpO
function
pulse
C-Lock
(ECG
synchr.)
mode
06-12-98
11:59
Select a menu position. Please confirm !
00:00
Change the settings. Please confirm again !
Standby / Configuration
CO
sample rate
ml/min
amplitude
real-
mmHg CO
units CO
2
insp. CO
2
sidestream
O
-measurem.
2
06-12-98
11:59
Select a menu position. Please confirm !
00:00
Change the settings. Please confirm again !
Standby / Configuration
pEEG
SEF / SMF / EMG
SEF / δ.Ratio / EMG
L.SEF / R.SEF / EMG
SMF / δ.Ratio /EMG
SEF / δ.Ratio / δ/θ/α/β
L.SEF / R.SEF / SMF
Amplitude
06-12-98
11:59
Select a menu position. Please confirm !
00:00
Change the settings. Please confirm again !
Settings
ECG
mode
adult
/Pleth
2
NiBP
alarm limits
on
off
iBP locations
parameter
on
off
iBP channels
screen
SpO
/ Pleth
acoustic
2
on
off
CO
/O
2
2
list entry
slow
normal
pEEG
transport
fast
RS 232 (Medibus)
record (printer)
basic config.
Settings
ECG
mode
adult
/O
2
2
NiBP
alarm limits
60
200
iBP locations
parameter
iBP channels
screen
2
0-75
0-110
SpO
/ Pleth
acoustic
2
mmHg kPa
CO
/O
2
2
list entry
Vol.%
pEEG
on
off
transport
on
off
RS 232 (Medibus)
record (printer)
basic config.
Settings
ECG
mode
adult
NiBP
alarm limits
iBP locations
parameter
iBP channels
screen
SpO
/ Pleth
acoustic
2
CO
/O
2
2
list entry
pEEG
transport
RS 232 (Medibus)
record (printer)
basic config.
alarms inactive !
neon.
alarms inactive !
neon.
alarms inactive !
neon.