Special Features Of Alarm Signalling - Dräger Cicero EM Instructions For Use Manual

Integrated anaesthetic workstation
Table of Contents

Advertisement

System screen functions
Alarm concept
Special alarm features
Related parameters are combined so that the
user is not disturbed by unnecessary cascade
alarms.
Combination of Apnea alarms from pressure,
flow and CO
measurement:
2
Measurement
Result
CO
No breathing
2
Flow
Breathing detected
Pressure
Breathing detected
CO
No breathing
2
Flow
No breathing
Pressure
Breathing detected
CO
No breathing
2
Flow
Breathing detected
Pressure
No breathing
CO
Breathing detected
2
Flow
No breathing
Pressure
No breathing
CO
Breathing detected
2
Flow
Breathing detected
Pressure
No breathing
CO
Breathing detected
2
Flow
No breathing
Pressure
Breathing detected
If a parameter fails for obvious reasons, the combination
is broken up and a separate alarm is signalled for each
parameter. Only the CO
2
»MAN/SPONT« mode.
82
Message
APNEA CO2 !!!
APNEA !!!
APNEA !!!
APNEA !!!
CHECK PRESSURE !
FLOW SENSOR !
Apnea alarm is signalled in
Special features
With indication of the time elapsed
in seconds
With indication of the time elapsed
in seconds
With indication of the time elapsed
in seconds

Advertisement

Table of Contents
loading