Dräger Cicero EM Instructions For Use Manual page 129

Integrated anaesthetic workstation
Table of Contents

Advertisement

Connecting the anaesthetic gas scavenging system
Hook the receiving system to the mount: position the
slots of the receiving system over the two holders and
slide down.
Connecting the transfer hoses
1 Pass the transfer hose from behind through the hole
in the Cicero EM and connect the hose to the scaven-
ging connector of the breathing system.
2 Connect the transfer hose to the connector on the
receiving system.
Make sure the connection is secure!
3 Make sure that the connector for the second transfer
hose is sealed with a screw plug.
Do not seal the openings of the receiving system.
Otherwise the breathing system may be drained!
Connect sample line
for sampling CO
/ O
and anaesthetic agent.
2
2
If a Y-piece with Luer lock is used:
4 Connect the sample line to the Luer lock of the
Y-piece.
5 Connect the sample line to the Luer lock of the
water trap.
If a Y-piece without Luer lock is used:
6 Insert T-piece with filter into the patient connection
of the Y-piece – filter facing upwards to prevent
congestion by droplets of liquid.
7 Connect the sample line to the Luer lock of the
filter in the T-piece.
5 Connect the sample line to the Luer lock of the water
trap.
Only use original sample lines, others may lead to
changes in the machine's technical specifications!
D
1
D
3
2
1
5
FULL
7
6
Reassembling machine
Reassembling the parts
4
129