General Settings; Acoustic; List Entry - Dräger Cicero EM Instructions For Use Manual

Integrated anaesthetic workstation
Table of Contents

Advertisement

Configuring the general settings
The following can be switched on (»yes«) or off (»no«):
units
Indication of the units.
Temperature
display
Option to display the body temperatures in
the bottom right-hand corner of the screen
(below the softkeys).
This display configuration saves space for
a numerical module.

Acoustic

The volume and tones are defined here.
»0« equals »off«, while »1« is the lowest and »9« the
highest volume.
alarm sound
Volume of the alarm
pulse tone
Volume of the pulse tone
SpO
-modulated
2
ECG pulse tone
The ECG pulse tone can optionally
be modulated with the SpO
tone sequence
Specifies the tone sequence.
Either ISO or Dräger standard.

List entry

A menu is displayed to allow the user to select at
what times (fixed intervals) or events (warning, caution
message, NiBP measurement) list entries are to be
made.
In addition, this menu defines the source of the blood
pressure value and whether EEG entries are required.
If EEG entries are selected, they replace the temperature
entries.
Standby / Configuration
unit
temperature-
display
06-12-98
11:59
00:00
Standby / Configuration
alarm sound
pulse tone
pulse tone
source
tone sequence
value.
2
06-12-98
11:59
00:00
Standby / Configuration
06-12-98
11:59
00:00
System screen functions
Configuring the system screen in »Standby«
Settings
mode
adult
alarm limits
parameter
screen
standard - screen 1
acoustic
list entry
standard - screen 2
standard - screen 3
transport
colour
RS 232 (Medibus)
record (printer)
curve speed
basic config.
yes
no

general settings

yes
no
Screen configuration menu. Please confirm !
Settings
mode
adult
alarm limits
parameter
screen
1 2 3 4 5 6 7 8 9
acoustic
0 1 2 3 4 5 6 7 8 9
list entry
ECG
Pleth.
transport
DRÄGER
Euro-Stand.
RS 232 (Medibus)
record (printer)
basic config.
Alarm sound and pulse tone configuration menu. Please confirm !
Settings
mode
adult
alarm limits
parameter
screen
acoustic
time interval
1 2 5 10 off
trigg. by warning
on
off
list entry
trigg. by caution
on
off
transport
with every
BP source
iBP 1 iBP 2
NIBP
RS 232 (Medibus)
pEEG
on
off
record (printer)
basic config.
List configuration menu. Please confirm !
alarms inactive !
neon.
alarms inactive !
neon.
alarms inactive !
neon.
61