Then switch to »Standby« and call up the leakage test
T
(see page 46) so that the new compliance can be
calculated:
Invoking the leakage test
1 Press
on the ventilator for at least three
seconds or press the softkey »Leakage test« on the
system screen -
2 Display in dialogue field on ventilator:
IPPV leakage test followed by: IPPV leak = xx mL
The menu for the leakage test appears on the system
screen followed by the current leakage and
compliance values a few seconds later.
The manual ventilation bag and its hose are not
included in the test!
The system switches back to »Standby« after the test.
The system screen should be switched over to »Neon«
mode (see page 49), in order to adapt the NIBP starting
pressure and the alarm limits to the patient.

Anaesthesia ventilation with the Kuhn system

M
Prepare the Kuhn set in accordance with separate
Instructions for Use -
Connect anaesthetic gas scavenging hose.
3 Press
on ventilator for at least one second -
4 Set pressure limiting valve APL to »MAN« -
Connect inspiration hose via
5 connecting sleeve -
6 Expiration connector remains open - Symbol:
7 Connect concertina hose and
breathing bag to manual ventilation
connector.
The pressure indicated on the system screen is not
identical with the actual airway pressure. Reason: higher
flow resistance of the fresh gas hose in the Kuhn system.
The higher the fresh gas flow, the greater the difference.
In case the Cicero EM is equipped with an external fresh
gas outlet (see page 29) this should be used.
National regulations on the minimum monitoring
requirement must be observed!
The breathing gas can be monitored by
8 the measuring connection on the mask manifold or
5 at the inspiration output (switch the system screen to
HLM mode for this purpose)
4
Symbol:
Symbol:
8
MAN
SPONT
IPPV
IPPV Leakage test
2
P
max
T
IP
:T
I
SIMV
PCV
V
T
f
T
:T
PEEP
IPPV
I
E
P
PCV
Flow
O
AIR
2
3
MAN
SPONT
IPPV
Lecktest ?
P
max
T
IP
:T
I
SIMV
PCV
V
f
T
:T
T
IPPV
I
E
P
PCV
Flow
D
5
6
7
Anaesthesia ventilation
Paediatric use
1
TEST
O
f
IMV
TEST
PEEP
f
IMV
39

Advertisement

Table of Contents
loading

Table of Contents