Stripping Down Machine - Dräger Cicero EM Instructions For Use Manual

Integrated anaesthetic workstation
Table of Contents

Advertisement

Care

Stripping down machine

Care
Stripping down machine
Swing up table top to remove piston pump:
(only possible if the device was previously in »Standby«)
1 Swing lever as far as possible to unlock.
2 Hold breathing system by handle and lift out.
Do not damage flat seals - avoid contact with sharp
edges. It must be put down so that the sealing
elements are not subject to constant pressure.
4 Caution!
Temperature of heater for breathing
system equals approx. 60 °C -
3 Handle pointing up - lift out piston pump.
Do not damage flat seals - avoid contact with sharp
edges. It must be put down so that the sealing
elements are not subject to constant pressure.
1 Swing lever back to original position, otherwise table
top cannot be reclosed.
Remove anaesthetic gas scavenging system
Refer to separate Instructions for Use for anaesthetic
gas scavenging system.
Unplug anaesthetic gas scavenging hose from wall
socket -
and remove from the anaesthetic gas scavenging
system.
5 Remove the transfer hose between the breathing
system and the anaesthetic gas scavenging system.
To do so:
Press button to unlock hose connector and pull
downwards. Hose remains on connector.
The anaesthetic gas scavenging system can be
removed from the back panel of the Cicero EM as a
single unit.